Wireless communication and wireless power transfer system for implantable medical device

Zhang Zhang, Chao Chen, Tairan Fei, Hao Xiao, Guangjun Xie and Xin Cheng

+ Author Affiliations

 Corresponding author: Xin Cheng, Email: ceciliacheng1017@163.com

PDF

Turn off MathJax

Abstract: Traditional magnetically coupled resonant wireless power transfer technology uses fixed distances between coils for research, to prevent fluctuations in the receiving voltage, and lead to reduce transmission efficiency. This paper proposes a closed-loop control wireless communication wireless power transfer system with a wearable four-coil structure to stabilize the receiving voltage fluctuation caused by changes in the displacement between the coils. Test results show that the system can provide stable receiving voltage, no matter how the distance between the transmitting coil and the receiving coil is changed. When the transmission distance is 20 mm, the power transfer efficiency of the system can reach 18.5% under the open-loop state, and the stimulus parameters such as the stimulation period and pulse width can be adjusted in real time through the personal computer terminal.

Key words: implantable medical equipmentfour-coil insertion resonance structurewireless communicationwireless power transfer



[1]
Xue R F, Cheng K W, Jie M. High-efficiency wireless power transfer for biomedical implants by optimal resonant load transformation. Circuits Syst I, 2013, 60, 867 doi: 10.1109/TCSI.2012.2209297
[2]
Sun T, Xie X, Wang Z. Wireless power transfer for medical microsystems. Berlin: Springer, 2013
[3]
Liu C H, Jiang C Q, Song J G, et al. An efficiency sandwiched wireless power transfer system for charging implantable cardiac pacemaker. IEEE Trans Ind Electron, 2019, 66(5), 4108 doi: 10.1109/TIE.2017.2840522
[4]
Kurs A, Karalis A, Moffatt R, et al. Wireless power transfer via strongly coupled magnetic resonances. Science, 2007, 317(5834), 83 doi: 10.1126/science.1143254
[5]
O’Driscoll S, Poon A S Y. Meng T H. A mm-sized implantable power receiver with adaptive link compensation. IEEE International Solid-State Circuits Conference, 2009, 294
[6]
Lee B, Kiani M, Ghovanloo M. A triple-loop inductive power transmission system for biomedical applications. IEEE Trans Biomed Circuits, 2016, 10, 138 doi: 10.1109/TBCAS.2014.2376965
[7]
Zhou W S, Sandeep S, Wu P D, et al. A wideband strongly coupled magnetic resonance wireless power transfer system and its circuit analysis. IEEE Microwave Wireless Compon Lett, 2018, 28(12), 1152 doi: 10.1109/LMWC.2018.2876767
[8]
Zhao Y L, Tang X, Wang Z H, et al. An inductive power transfer system with adjustable compensation network for implantable medical devices. IEEE Asia Pacific Conference on Circuits and Systems, 2019
[9]
Liu L, He L, Zhang Y, et al. A battery-less portable ECG monitoring system with wired audio transmission. IEEE Trans Biomed Circuits Syst, 2019, 13(4), 697 doi: 10.1109/TBCAS.2019.2923423
[10]
Mury T, Fusco V F. Series-L/parallel-tuned comparison with shunt-C/series-tuned class-E power amplifier. IEE Proc-Circuits, Devices Syst, 2005, 152(6), 709 doi: 10.1049/ip-cds:20050026
[11]
Kumar A, Mirabbasi S, Chiao M. Resonance-based wireless power delivery for implantable devices. Proc IEEE Biomedical Circuits and Systems Conf, 2009, 25
[12]
Qian L B, Sang J F, Xia Y S, et al. Investigating on through glass via based RF passives for 3-D integration. IEEE J Electron Devices Soc, 2018, 6, 755 doi: 10.1109/JEDS.2018.2849393
[13]
Xu H, Bihr U, Becker J, et al. A multi-channel neural stimulator with resonance compensated inductive receiver and closed-loop smart power management. IEEE International Symposium on Circuits and Systems, 2013, 638
[14]
Seo D W, Khang S T, Chae S C, et al. Open-loop self-adaptive wireless power transfer system for medical implants. Microwave Opt Technol Lett, 2016, 58(6), 1271 doi: 10.1002/mop.29796
Fig. 1.  Schematic diagram of the system.

Fig. 2.  The block diagram of the WC-WPT system.

Fig. 3.  Inverse class E PA.

Fig. 4.  The structure of transmitting coil and repeating coils.

Fig. 5.  The PSSP type four-coil resonant topology.

Fig. 6.  (Color online) The relationship between efficiency and coupling coefficient: (a) k12, (b) k23 and (c) k34.

Fig. 7.  The specific adjustment process.

Fig. 8.  (Color online) The GUI for control of stimulator parameters.

Fig. 9.  (Color online) The experimental system.

Fig. 10.  (Color online) Performance of PTE: (a) two-coil Structure, and (b) four-coil structure.

Fig. 11.  (Color online) Variation of receiving voltage at (a) open-loop system and (b) closed-loop system.

Table 1.   The calculation formula of main components of the inverter circuit.

ComponentLLfCxRx
Computing formula${ {{L} } = \dfrac{ { {{V} }_{ {\rm{DC} } }^2} }{ { { {\pi \omega } }{ {{P} }_0} } } }$${ { {{L} }_{\rm{f} } } = \dfrac{ {\left( {2{ { {\pi } }^2} + 8} \right){ {{R} }_{\rm{x} } } }}{ {4{ {{f} }_0} } } }$${ { {{C} }_{\rm{x} } } = \dfrac{ { { {\pi } }\left( { { { {\pi } }^2} - 4} \right){ {{P} }_0} } }{ {2\left( { { { {\pi } }^2} + 4} \right){ {\omega V} }_{ {\rm{DC} } }^2} } }$${ { {{R} }_{\rm{x} } } = \dfrac{ {\left( { { { {\pi } }^2} + 4} \right){{V} }_{ {\rm{DC} } }^2} }{ {8{ {{P} }_0} } } }$
DownLoad: CSV

Table 2.   The parameters of the coils and resonant compensation capacitance at 1 MHz.

Coil numberDiameter (cm)Inductance (μH)Parasitic resistance (Ω)Compensation capacitance (nF)
Coil 16. 012. 030. 222. 11
Coil 28. 118. 580. 261. 36
Coil 310. 022. 620. 271. 12
DownLoad: CSV

Table 3.   Benchmarking of WPT system.

Publication2013[13]2016[6]2016[14]2018[7]This work
Power carrier frequency (MHz)13.813.5613.56100.11
Coil configuration23244
Closed-loop controlYesYesNoNoYes
Real-time controlNoNoYesNoYes
Wireless communicationLSKLSK//nRF24L01
PTE/13.5% @ 20 mm10% @ 20 mm12.5% @ 400 mm18.5% @ 20 mm
Tx coil diameter (mm)/282080120
Rx coil diameter (mm)/341080 21
Potential applicationIMDsIMDsIMDs/IMDs
DownLoad: CSV
[1]
Xue R F, Cheng K W, Jie M. High-efficiency wireless power transfer for biomedical implants by optimal resonant load transformation. Circuits Syst I, 2013, 60, 867 doi: 10.1109/TCSI.2012.2209297
[2]
Sun T, Xie X, Wang Z. Wireless power transfer for medical microsystems. Berlin: Springer, 2013
[3]
Liu C H, Jiang C Q, Song J G, et al. An efficiency sandwiched wireless power transfer system for charging implantable cardiac pacemaker. IEEE Trans Ind Electron, 2019, 66(5), 4108 doi: 10.1109/TIE.2017.2840522
[4]
Kurs A, Karalis A, Moffatt R, et al. Wireless power transfer via strongly coupled magnetic resonances. Science, 2007, 317(5834), 83 doi: 10.1126/science.1143254
[5]
O’Driscoll S, Poon A S Y. Meng T H. A mm-sized implantable power receiver with adaptive link compensation. IEEE International Solid-State Circuits Conference, 2009, 294
[6]
Lee B, Kiani M, Ghovanloo M. A triple-loop inductive power transmission system for biomedical applications. IEEE Trans Biomed Circuits, 2016, 10, 138 doi: 10.1109/TBCAS.2014.2376965
[7]
Zhou W S, Sandeep S, Wu P D, et al. A wideband strongly coupled magnetic resonance wireless power transfer system and its circuit analysis. IEEE Microwave Wireless Compon Lett, 2018, 28(12), 1152 doi: 10.1109/LMWC.2018.2876767
[8]
Zhao Y L, Tang X, Wang Z H, et al. An inductive power transfer system with adjustable compensation network for implantable medical devices. IEEE Asia Pacific Conference on Circuits and Systems, 2019
[9]
Liu L, He L, Zhang Y, et al. A battery-less portable ECG monitoring system with wired audio transmission. IEEE Trans Biomed Circuits Syst, 2019, 13(4), 697 doi: 10.1109/TBCAS.2019.2923423
[10]
Mury T, Fusco V F. Series-L/parallel-tuned comparison with shunt-C/series-tuned class-E power amplifier. IEE Proc-Circuits, Devices Syst, 2005, 152(6), 709 doi: 10.1049/ip-cds:20050026
[11]
Kumar A, Mirabbasi S, Chiao M. Resonance-based wireless power delivery for implantable devices. Proc IEEE Biomedical Circuits and Systems Conf, 2009, 25
[12]
Qian L B, Sang J F, Xia Y S, et al. Investigating on through glass via based RF passives for 3-D integration. IEEE J Electron Devices Soc, 2018, 6, 755 doi: 10.1109/JEDS.2018.2849393
[13]
Xu H, Bihr U, Becker J, et al. A multi-channel neural stimulator with resonance compensated inductive receiver and closed-loop smart power management. IEEE International Symposium on Circuits and Systems, 2013, 638
[14]
Seo D W, Khang S T, Chae S C, et al. Open-loop self-adaptive wireless power transfer system for medical implants. Microwave Opt Technol Lett, 2016, 58(6), 1271 doi: 10.1002/mop.29796
  • Search

    Advanced Search >>

    GET CITATION

    shu

    Export: BibTex EndNote

    Article Metrics

    Article views: 3811 Times PDF downloads: 109 Times Cited by: 0 Times

    History

    Received: 17 March 2020 Revised: 06 April 2020 Online: Accepted Manuscript: 02 June 2020Uncorrected proof: 03 June 2020Published: 01 October 2020

    Catalog

      Email This Article

      User name:
      Email:*请输入正确邮箱
      Code:*验证码错误
      Zhang Zhang, Chao Chen, Tairan Fei, Hao Xiao, Guangjun Xie, Xin Cheng. Wireless communication and wireless power transfer system for implantable medical device[J]. Journal of Semiconductors, 2020, 41(10): 102403. doi: 10.1088/1674-4926/41/10/102403 Z Zhang, C Chen, T R Fei, H Xiao, G J Xie, X Cheng, Wireless communication and wireless power transfer system for implantable medical device[J]. J. Semicond., 2020, 41(10): 102403. doi: 10.1088/1674-4926/41/10/102403.Export: BibTex EndNote
      Citation:
      Zhang Zhang, Chao Chen, Tairan Fei, Hao Xiao, Guangjun Xie, Xin Cheng. Wireless communication and wireless power transfer system for implantable medical device[J]. Journal of Semiconductors, 2020, 41(10): 102403. doi: 10.1088/1674-4926/41/10/102403

      Z Zhang, C Chen, T R Fei, H Xiao, G J Xie, X Cheng, Wireless communication and wireless power transfer system for implantable medical device[J]. J. Semicond., 2020, 41(10): 102403. doi: 10.1088/1674-4926/41/10/102403.
      Export: BibTex EndNote

      Wireless communication and wireless power transfer system for implantable medical device

      doi: 10.1088/1674-4926/41/10/102403
      More Information
      • Corresponding author: Email: ceciliacheng1017@163.com
      • Received Date: 2020-03-17
      • Revised Date: 2020-04-06
      • Published Date: 2020-10-04

      Catalog

        /

        DownLoad:  Full-Size Img  PowerPoint
        Return
        Return